Molecular Devices SpectraMax 340 PC 96 & 384 Microplate Reader w WARRANTY
Molecular Devices SpectraMax 340 PC 96 & 384 Microplate Reader w WARRANTY
$2,750 (USD)
Description
Molecular Devices SpectraMax 340 PC 96 & 384 Microplate Reader w WARRANTY
This is a Nice Used Molecular Devices SpectraMax 340PC 384 Microplate Spectrophotometer. Unit is in Overall Excellent Condition. Reader Powers Up and Passes All Start Up Diagnostics and Proceeds to the Main Screen. The Unit Has Been Tested Against a Calibration Plate at Various Wavelegths and the Results are Accurate within 5% of Target and Provided Below. Unit is Being Offered as a Complete System That Will Be Ready For Use Upon Receipt. Please Note Pictured Calibration Plate is Not Included. Unit Will Guaranteed 30 Days Allowing Enough Time for Your Lab To Fully Test and Inspect. Please Add $75 Flat Rate Ground Shipping with the US.
Included:
- SpectraMax 340PC 384 Reader SN# LNR 02006
- Dell Laptop with Spectramax 5.4 Installed
- All Power and Communications Cables
- User's Manual and SoftMax Pro Manual Loaded on Computer
Specifications:
- Wavelength Range: 340-850 nm
- Wavelength Selection: Monochromator, tunable in 1-nm increments
- Photometric Range: 0 to 4.000 OD
- Wavelength Bandwidth: 2 nm
- Endpoint Read Time: 9 seconds
- Temperature Range: Ambient to 45 C
- Programmable Shaking
- Kinetic Read Time: 9-second minimum interval
This Reader's Calibration Test Plate Results:
1) 405nm Reading: * Expected .679, 1.214, 1.815 * Actual: .686, 1.229, 1.836
2) 450nm Reading: * Expected .627, 1.113, 1.666 * Actual: .628, 1.117, 1.670
3) 490nm Reading: * Expected .626, 1.101, 1.646 * Actual: .630, 1.108, 1.659
4) 550nm Reading: * Expected .619, 1.077, 1.589 * Actual: .624, 1.087, 1.598
5) 630nm Reading: * Expected .614, 1.069, 1.601 * Actual: .621, 1.077, 1.612
6) 750nm Reading: * Expected .474, 0.782, 1.166 * Actual: .479, 0.785, 1.173
Specifications
Manufacturer | Molecular Devices |
Model | Spectramax 340 PC 384 |
Condition | Used |
Stock Number | 14666889 |
* Wavelength Range | 340-850 nm |
* Photometric Range | 0 to 4.000 OD |
* Wavelength Bandwidth | 2 nm |
* Endpoint Read Time | 9 seconds |
* Temperature Range | Ambient to 45 C |
* Kinetic Read Time | 9-second minimum interval |